Mapping the field uniformity of a permanent dipole in a 12 mm gap
CORIA, a research laboratory in France working on fluid dynamics and plasma physics, needed to characterise how uniform the magnetic field of a permanent dipole is inside its 12 mm gap. They mapped the field with the 3-axis Hall probe of a SENIS 3MH6-E teslameter moved through the gap on a manual 3D translation stage.
The task
The dipole consists of two magnets in a frame with opposite poles facing each other across a 12 mm gap (sketch). CORIA wanted to know how uniform the field is in the gap.

The challenge
The gap is only 12 mm wide, so the probe has to fit inside it.
How it was measured
- The teslameter’s 3-axis Hall probe measures the three field components Bx, By and Bz.
- The probe was mounted on a manual 3D translation stage and moved through a volume in the centre of the gap.
- The teslameter’s data acquisition functions were used to make the mapping as precise and efficient as possible.
Result
The compact probe fitted easily within the 12 mm gap and gave high spatial resolution. The map shows the total field |B| in the centre of the gap, on a colour scale from 0.96 T to 1.00 T.

“[The SENIS teslameter] allows us to perform fast, accurate and complete mappings of the magnetic fields generated by different sources involved in our experiments in fluid dynamics and plasma physics.” Paul Dubois, PhD Student, CORIA, France
Instrument used
3MH6-E high-precision teslameter with a 3-axis Hall probe.
Have a similar measurement task?
Describe the field source, the space available for the probe and the field values you need to measure, and contact SENIS.